Observation Possible Cause Action
Too much sample was left in
loading blade.
Ensure that you are loading only
14.5 µL of reaction mix into the
loading blade.
Chapter 5 Troubleshooting
Troubleshooting chip images using the Chip View
5
QuantStudio
®
3D Digital PCR System User Guide
85